Combine integrated vacuum filtration, mixing, and temperature control in one system to automate even complex applications. The epMotion 5075vt allows true walk-away automation of purification protocols and other assays that require an Eppendorf ThermoMixer. The 5075vt is controlled by MultiCon industry PC and can be configured with CleanCap and up to 2 thermal modules. More information

Order Info

Product Information

The epMotion 5075vt is all about productivity and flexibility. With its integrated vacuum station and Eppendorf ThermoMixer® the epMotion 5075vt allows true walk-away automation of purification protocols that rely on vacuum filtration or magnetic beads.

Integrated Vacuum Station


Loaded and unloaded with the gripper, the vacuum station adapts automatically to any filter plate type. Silent operation of internal pump with no need for extra tubes, wiring, or reservoir maintenance. The Eppendorf silicone mat accessory helps to save time and money
by allowing you to use only portions of a 96-well plate when working with smaller sample numbers. The VAC thermo lid accessory also reduces the extraction time substantially
by accelerating the ethanol evaporation step.

epMotion® 5075vt

Eppendorf ThermoMixer®


Fully integrated Eppendorf ThermoMixer with 2D-Mix-Control improves yield and application results, e.g. resuspending bacteria pellets or magnetic beads. Peltier element for heating and cooling of sample and/or reagents. The following epMotion versions are equipped with Eppendorf ThermoMixer capability, M5073, 5075t, 5075m, 5075vt. A Eppendorf ThermoMixer upgrade product is available for epMotion 5075l, 5075v, and successor versions 5075LH and 5075VAC.

epMotion® 5075vt

Reuse Tips Feature


The new epMotion systems can reuse tips as e.g. economical choice for multiple wash steps
of the same well. Here special epT.I.P.S. Motion SafeRack tips are recommended. Pipette tips will be placed back into the position of the box they were removed from. Returned tips will be reused within an application for the same source vessel.

› Individual compartments in rack to avoid cross-contamination between used tips
› Available as standard and PCR clean quality for 50, 300 and 1000 μL

epMotion® 5075vt

Features

  • Same as 5075v, plus
  • Integrated Eppendorf Thermomixer® with 2DMix-Control technology
  • Temperature control range 15 °C below RT to 95 °C
  • Automatic fixing of labware for mixing
  • Pipetting at standard positions during shaking

Applications

  • Like epMotion 5075v, plus
  • Bead applications with mixing and tempering
  • Sequencing and PCR clean-up
  • Nucleic acid purification
  • Cell disruption
  • Immunoassays
 epMotion® 5075vt
Gripper carrying capacity1,200 g
Mixing frequencyOff, 300 – 2,000 rpm, optimized for labware and application
OperationElectronic
Pipetting typeAir-cushion
SLAS/ANSI deck positions12
Temperature homogeneity≤ 10 % on the block
Heating rate5 °C/min, on the block
Cooling rate3 °C/min, on the block, above RT
Power supply100 – 240 V ±10 %, 50 – 60 Hz ±5 %
Max. power consumption700 W
Dimensions (W × D × H)107 × 61 × 67 cm / 43 × 24 × 27 in
Weight w/o accessories90 kg / 198.4 lb
Temperature range–15 °C below RT to 95 °C measured in labware at 75 % filling height
Thermal module (optional) 
Cooling rate4 °C/min
Heating rate9 °C/min
Temperature range0 °C - 110 °C
ThermoMixer 
Max. load1,000 g (2.2 lb)
Mixing period5 s – 120 min
Speed300 rpm – 2,000 rpm
Temperature range15 °C below RT to 95 °C
Vacuum unit 
Suction range0.1-85 kPa
Suction time1 sec to 36 min
Vacuum max. output35 NL/min
Conductor – X, Y, Z Positioning 
Random measurement error±0.1 mm
Systematic measurement error±0.3 mm
Detector 
Optical confocal infrared detectorContact-free recognition of fill level, inserted tools, labware, tip types and quantities
Optical sensorContact-free recognition of fill level, inserted tools, labware, tip types and quantities
Dispensing tool 
Random measurement error (1 µL)1,2≤5 %
Random measurement error (1,000 µL)1,2≤0.15 %
Random measurement error (50 µL)1,2≤0.4 %
Systematic measurement error (1 µL)1,2±15 %
Systematic measurement error (1,000 µL)1,2±0.7 %
Systematic measurement error (50 µL)1,2±1.2 %
1in pipetting mode, free jet, without pre-wetting, with distilled water, at 20 °C
2See Eppendorf Application Note 168 for typical pipetting performance