The epMotion 5075 systems are larger than 5070 and 5073 systems.
With 15 worktable positions and more features, like gripper transport,
4 positions for exchangeable pipetting tools, the epMotion 5075 versions
have a higher application flexibility.
The epMotion 5075l is controlled by MultiCon industry PC and can be configured with CleanCap and up to 3 thermal modules. More information

Order Info

Product Information

The epMotion 5075l is the perfect solution for more complex pipetting tasks. It offers the same precision and accuracy as the epMotion 5070, making it the ideal tool for demanding applications such as the purification of nucleic acids with magnetic beads or routine applications with several plates. The 15 positions, option of automatically changing pipetting tools, and the plate gripper make the epMotion 5075l one of the most flexible automated liquid handling systems.

Large Deck Space


With up to 15 positions in SBS/ANSI format the epMotion 5075l gives
you all the flexibility you need for your application.

epMotion® 5075l

Gripper Tool


The gripper tool allows transport of labware on the worktable of epMotion 5073 or 5075 versions. It is loaded by automatic tool exchange and can be programmed easily in the software. The stacking of up to 5 plates is possible.

epMotion® 5075l

CleanCap - UV/HEPA


The CleanCap option can be configured for new epMotion 5073 & 5075 versions.
Here a UV light (254 nm) is used for decontamination and the HEPA filters provide
clean air inside of the epMotion.

epMotion® 5075l

Features

  • 15 ANSI/SLAS worktable positions
  • Maximum pipetting accuracy from 0.2 to 1,000 µL
  • Automatic exchange of all tools (gripper, 1-channel and 8-channel)
  • Optical sensor1) for detecting liquids, labware, tips
  • Compatible with tubes (0.2 mL to 50 mL) and microplates with up to 384 wells
  • Eppendorf MultiCon PC controller to run preinstalled LH assistant or epBlue software by touch or mouse
  • Optional software extensions for handling samples with barcodes or for integration to LIMS and ELN
  • Optional software extensions for integration to regulated process environments (GLP, GMP, 21 CFR)
  • CleanCap option – UV lamp and HEPA filter for decontamination and clean air conditions
  • 3 thermal module options for heating or cooling (0 – 110 °C) of samples or reagents

1) Patent US 6,819,437 B2

Applications

  • NGS library preparation
  • Distributing reagents and serial dilutions
  • Sample transfer from individual tubes to plates
  • Reformatting plates
  • Media change and other cell culture applications
  • Sample pooling
  • Real-time PCR set-up
  • Cell-based assay set-up
  • Immunoassay set-up
  • Setting concentrations or volumes
 epMotion® 5075l
Gripper carrying capacity1,200 g
OperationElectronic
Pipetting typeAir-cushion
Random measurement error (1 µL)≤5 %
SLAS/ANSI deck positions15
Systematic measurement error (1 µL)±15 %
Power supply100 – 240 V ±10 %, 50 – 60 Hz ±5 %
Max. power consumption700 W
Dimensions (W × D × H)107 × 61 × 67 cm / 43 × 24 × 27 in
Weight w/o accessories85 kg / 187.4 lb
Thermal module (optional) 
Cooling rate4 °C/min
Heating rate9 °C/min
Temperature range0 °C – 110 °C
Conductor – X, Y, Z Positioning 
Random measurement error±0.1 mm
Systematic measurement error±0.3 mm
Detector 
Optical confocal infrared detectorContact-free recognition of fill level, inserted tools, labware, tip types and quantities
Optical sensorContact-free recognition of fill level, inserted tools, labware, tip types and quantities
Dispensing tool 
Random measurement error (1 µL)1,2≤3 %
Random measurement error (1,000 µL)1,2≤0.15 %
Random measurement error (50 µL)1,2≤0.4 %
Systematic measurement error (1 µL)1,2±3 %
Systematic measurement error (1,000 µL)1,2±0.7 %
Systematic measurement error (50 µL)1,2±1.2 %
1in pipetting mode, free jet, without pre-wetting, with distilled water, at 20 °C
2See Eppendorf Application Note 168 for typical pipetting performance